Düven, GamzeÇetin, BarbarosKurtuldu, HüseyinGündüz, Gülten TiryakiTavman, ŞebnemKışla, Duygu2020-07-252020-07-2520191387-21761572-8781https://doi.org/10.1007/s10544-019-0407-8https://hdl.handle.net/11147/9549Microfluidics systems that have been emerged in the last 20years and used for processing the fluid in a microchannel structure at microliter levels are alternative to the conventional methods. The objective of the study is to develop a microfluidic platform for determination of the microbial load and the number of somatic cells in milk. For this purpose, a polydimethylsiloxane (PDMS) chip with a channel size of 300mx60m was produced. Cells/bacteria labeled with fluorescent stain in milk were counted with the proposed microfluidic platform and the results were compared with the reference cell concentration/the bacterial counts by conventional method. It was found that our platform could count somatic and bacterial cells with an accuracy above 80% in 20min run for each analysis. The portable overall platform has an overall dimension of 25x25x25 cm and weighs approximately 9kg.eninfo:eu-repo/semantics/openAccessMicrofluidicsQuality of milkSomatic cell countBacteria countA Portable Microfluidic Platform for Rapid Determination of Microbial Load and Somatic Cell Count in MilkArticle2-s2.0-8506736472010.1007/s10544-019-0407-810.1007/s10544-019-0407-8