Please use this identifier to cite or link to this item: https://hdl.handle.net/11147/10076
Title: Thermal characterization of Ag and Ag plus N ion implanted ultra-high molecular weight polyethylene (UHMWPE)
Authors: Urkaç, Emel Sokullu
Oztarhan, A.
Tıhmınlıoğlu, Funda
Kaya, N.
Ila, D.
Muntele, C.
Tek, Z.
Keywords: UHMWPE
Ag
Ag plus N
ion implantation
DSC
TGA
SEM
Publisher: Elsevier Ltd.
Abstract: Most of total hip joints are composed of ultra-high molecular weight polyethylene (UHMWPE). However, as ultra-high molecular weight polyethylene is too stable in a body, wear debris may accumulate and cause biological response such as bone absorption and loosening of prosthesis. In this study, ultra-high molecular weight polyethylene samples were Ag and Ag + N hybrid ion implanted by using MEVVA ion implantation technique to improve its surface properties. Samples were implanted with a fluence of 10(17) ion/cm(2) and extraction voltage of 30 kV. Implanted and unimplanted samples were investigated by thermo-gravimetry analysis (TGA), differential scanning calorimetry (DSC), X-ray diffraction (XRD) analysis, scanning electron microscopy (SEM), optical microscopy (OM) and contact Angle measurement. Thermal characterization results showed that the ion bombardment induced an increase in the % crystallinity, onset and termination degradation temperatures of UHMWPE. (c) 2007 Elsevier B.V. All rights reserved.
Description: 19th International Conference on Application of Accelerators in Research and Industry
URI: https://doi.org/10.1016/j.nimb.2007.04.102
https://hdl.handle.net/10076
ISSN: 0168-583X
Appears in Collections:WoS İndeksli Yayınlar Koleksiyonu / WoS Indexed Publications Collection

Show full item record



CORE Recommender

SCOPUSTM   
Citations

18
checked on Apr 5, 2024

WEB OF SCIENCETM
Citations

14
checked on Mar 27, 2024

Page view(s)

84
checked on Apr 22, 2024

Google ScholarTM

Check




Altmetric


Items in GCRIS Repository are protected by copyright, with all rights reserved, unless otherwise indicated.